Name: | DOG-208F pure water dissolved oxygen electrode |
---|---|
Measuring range: | 0-100ug/L; 0-20mg/L |
Resolution: | 0.1µg/L |
Electrode material: | SS316 |
Response time: | ≤3min (90%, 20℃) |
Measuring range: 0-100ug/L; 0-20mg/L
Electrode material: SS316
Temperature compensation resistance: NTC: 2.252K, Pt100, Pt1000, etc.
Sensor life: >3 years
Cable length: standard 5m (double shielded)
Lower detection limit: 0.1 µg/L (ppb) (20°C)
Upper limit of measurement: 20 mg/L (ppm)
Response time: ≤3min (90%, 20℃)
Polarization time: >8h
Minimum flow rate: 5cm/s; 515L/h
Drift: <3%/month
Measurement error: <±1ppb
Output current: 20-25nA/0.01 mg/L Note: Maximum current 20-25uA
Polarization voltage: 0.7V
Zero oxygen: <5ppb(60min)
Calibration interval: >60 days
Working temperature: 0-60℃
1. DOG-208F dissolved oxygen electrode applies polarographic principle
2. Use platinum (Pt) as cathode and Ag/AgCl as anode
3. The electrolyte is 0.1M potassium chloride (KCl)
4. Use American imported silicone rubber permeable membrane as breathable membrane
5. It has silicone rubber and steel gauze
6. Collision resistance, corrosion resistance, high temperature resistance and non-deformation performance
It is used in places with trace oxygen content such as thermal power plants, power plants and power stations desalinated water, boiler feed water
It is used in places with trace oxygen content such as thermal power plants, power plants and power stations desalinated water, boiler feed water